Accuracy means the measured value is close to the true value (via standard-gas response-factor calibration); repeatability means multiple measurements under the same conditions are consistent (via programmatic automatic injection). The advantage of online systems is mainly in repeatability, but accuracy still needs manual calibration as a safeguard.
| Source | Manifestation | Countermeasure |
|---|---|---|
| Leakage | Systematically low yield | Positive-pressure leak test |
| Response-factor drift | Value shift | Periodic standard-gas re-check |
| Uncalibrated light intensity | Non-comparable rates | Sample-position calibration |
Free consultation: Always just short of solid data? Perfectlight turns the μGAS1001 online analysis and light-intensity calibration into an auditable QC workflow, making your data directly publishable.
Tel: 400-1161-365 / 010-62128921 Perfectlight Official Website
No. Accuracy relies on standard-gas calibration; repeatability relies on programmatic automatic injection; the online advantage is mainly in repeatability.
Include a blank and parallel sample per batch, record injection volume and column temperature, and use the PL-MW2000 to synchronously record light intensity for normalization.
Leakage, response-factor drift, and uncalibrated light intensity are the three main causes.