Match the detector to the gas, not to the budget. TCD covers H2, O2, N2, CO and CO2 at 10–100 ppm; FID covers CH4 and alcohols at 0.1–1 ppm but cannot see H2, CO2 or formaldehyde; PDHID reaches 10–100 ppb on every gas except He and Ne, at 2–4 times the instrument cost.
TCD compares the thermal conductivity of the column effluent against a reference flow of the same carrier gas. Response is proportional to the difference (ksample − kcarrier), so its sensitivity is set by the carrier you choose, not by the detector alone. At 25 °C, k is 0.187 W m−1 K−1 for H2, 0.151 for He, 0.0260 for N2, 0.0177 for Ar and 0.0166 for CO2.
FID pyrolyses the analyte in a hydrogen/air flame and counts CHO+ ions, so the response scales with carbon atoms and is essentially zero for fully oxidised carbon. CO2, CO, HCHO, HCOOH and H2O produce no signal at all — which is exactly the set of species that matters in CO2 reduction and in partial oxidation work.
PDHID uses a pulsed discharge in high-purity helium; the 13.5–17.7 eV photons ionise anything with an ionisation potential below that window. It is close to universal, which is both its value and its liability: the carrier itself must be clean enough not to be the signal.
| Detector | Gases it measures well | Typical LOD | Carrier | Effectively blind to | Relative cost |
|---|---|---|---|---|---|
| TCD | H2, O2, N2, CO, CO2, CH4 above 0.1 % | 10–100 ppm | He, Ar or N2 | ppb-level anything | 1.0 |
| FID | CH4, C2–C6, methanol, ethanol, acetone | 0.1–1 ppm | H2/air flame, N2 make-up | H2, CO, CO2, HCHO, HCOOH, H2O | 1.2–1.5 |
| FID + methaniser | CO and CO2 as CH4 | 0.05–0.5 ppm | H2 carrier, Ni catalyst at 350–400 °C | HCHO unless derivatised | 1.5–1.8 |
| PDHID | All permanent gases and C1–C4 except He and Ne | 10–100 ppb | He 99.9999 % plus getter | He, Ne | 2–4 |
Limits assume a 1 mL sample loop, a 30 m × 0.53 mm molsieve/porous-polymer column set and a signal-to-noise ratio of 3. Narrow-bore columns and larger loops buy a factor of 3–5; they cost run time and column life.
Hydrogen in helium is the classic mistake. The two conductivities differ by only 24 %, so the peak is small and easily buried in drift; switch to N2 or Ar and the same 1 % H2 sample produces a full-scale peak of the opposite polarity. Conversely CO2 in helium gives a large negative peak, which is fine once you set the polarity in the method rather than the software default.
Filament life is a carrier question too. A tungsten–rhenium filament run with H2 carrier and a trace of oxygen at 200 °C oxidises within weeks; the same filament in helium or nitrogen lasts years. If the method needs hydrogen as carrier for a methaniser, keep the TCD cell at the lowest temperature that still resolves the peak.
One ppm of any gas in a 100 mL headspace at 25 °C is 4.1 nmol, because 100 mL is 4.08 mmol of ideal gas. That single conversion decides most arguments. A rate of 0.2 µmol h−1 accumulated over one hour in 100 mL is 49 ppm — comfortably inside TCD range. A rate of 4 nmol h−1, which is what an apparent quantum yield near 0.01 % produces, is 1 ppm and sits exactly at the TCD floor; here PDHID or a 5 mL loop is not a luxury but the only way to see a peak at all.
For CO2 reduction the split is harsher. CH4 and methanol are FID species; CO needs a methaniser or a PDHID; CO2 itself must be measured on TCD or by difference, never on FID. Formaldehyde is invisible on both FID and TCD at room temperature and needs DNPH derivatisation with HPLC, or a dedicated electrochemical cell. A rig with only a TCD therefore cannot rule out formaldehyde as a product, and reviewers increasingly ask.
State the carrier and its purity grade, the column set and temperature programme, the detector temperature and, for PDHID, the discharge frequency and voltage. Give the calibration gas certificate concentrations and the range actually bracketed by the standards, not the cylinder's nominal value. Report the limit of detection as three times the standard deviation of ten blank injections, and say whether a methaniser was fitted. Four lines of this prevent most of the questions a gas-analysis figure attracts.
Not in a single channel with good sensitivity, because the carrier that suits one hurts the other. Helium gives a strong CO2 peak and a weak H2 peak; nitrogen or argon reverses that. Use a two-channel instrument with a molsieve channel on Ar for H2 and a porous-polymer channel on He for CO2, or accept 100 ppm limits on both.
Because FID only responds to reduced carbon. CO and CO2 give no signal, so a positive CO result requires a methaniser converting them to CH4 over a nickel catalyst at 350 to 400 °C. Formaldehyde and formic acid also fail to respond unless derivatised first.
Only below about 5 ppm, which corresponds to roughly 20 nmol in a 100 mL headspace. Above that a TCD on argon carrier reaches the same number with a simpler method and no ultra-high-purity helium supply. Buy PDHID when the science is trace products or isotope work, not for a routine screening rig.
Inject ten blanks and ten standards at twice the claimed limit, and require the standard peak to be at least three times the blank standard deviation with a retention time within 0.05 min. A limit quoted from a catalogue assumes a clean system; a real rig with a septum leak will not reproduce it.