Home > Technology > Instrument Knowledge > Lamp-to-Reactor Distance and Irradiance: The Inverse-Square Trap
2026-11-23109

Lamp-to-Reactor Distance and Irradiance: The Inverse-Square Trap

Quick answer

Halving the lamp-to-window distance does not quadruple the irradiance. On a typical 300 W xenon with a 60 mm reflector, 300 mm to 150 mm raises it about 3.1 times, not 4, because the reflector aperture and the 3 mm arc make the source finite. Report distance from the arc plus the reference plane.

What the inverse-square law assumes

E = I / d2 holds for a point source in vacuum with no reflector and no aperture. A xenon short-arc lamp violates all three: the arc is 2–4 mm long, the parabolic reflector turns a large area into a second emitter, and the condenser adds a virtual image of the arc behind the housing. The measured falloff therefore behaves like a point source located tens of millimetres behind the physical arc, not at it.

A simple offset model captures most of it. Replace d by (d + e), where e is the distance from the arc to the effective origin, typically 40–70 mm for a reflector-and-condenser assembly. With e = 60 mm, the ratio of irradiance at 150 mm to 200 mm is ((200+60)/(150+60))2 = 1.53, against 1.78 from the pure 1/d2 law. That 14 % gap is the difference between a clean Arrhenius fit and one with unexplained scatter.

Measured against ideal, normalised to 200 mm

Arc-to-window distanceIdeal 1/d2, relativeTypical measured, relativeCentre-to-edge spreadOn-axis, typical 300 W Xe
100 mm4.002.615–20 %350–450 mW/cm2
150 mm1.781.510–15 %180–250 mW/cm2
200 mm1.001.006–10 %100–150 mW/cm2
300 mm0.440.524–6 %50–75 mW/cm2
400 mm0.250.323–4 %30–45 mW/cm2
600 mm0.110.162–3 %15–22 mW/cm2

Values are for a 300 W xenon short-arc with a 60 mm parabolic reflector, unvignetted, measured on axis with a calibrated thermal sensor at 25 °C ambient; mW/cm2 scales with lamp power and with the reflector, so treat the absolute column as an order-of-magnitude check and the relative columns as the transferable part. The measured column comes from the (d + 60 mm) offset model and reproduces real benches to within about 10 %.

The reference plane is the number that matters

Two people can quote "200 mm from the lamp" and measure irradiance differing by 40 %. One means the housing face, the other the outer surface of the reactor window, and the 30–50 mm between those planes is worth more than 30 % at short distances. Fix the convention in writing: arc centre to the outside face of the window, with the window in place, at the reactor axis, with the beam perpendicular to within 2°.

Angle matters more than most people expect. A meter tilted 10° off normal reads about 1.5 % low; tilted 30° it reads 13 % low. A beam that hits the window at 45° loses another 8–10 % to Fresnel reflection on top of the cosine falloff. If the reactor sits off-axis to fit a support, the irradiance at the catalyst is not the irradiance on the bench.

Running a distance series that survives review

Vary one thing. Move the reactor, not the lamp, and re-measure irradiance at the window plane in every position. Do not interpolate from a single calibration or assume the 1/d2 ratio. Record the actual mW/cm2 beside each rate, so a reviewer can see whether the apparent order in light intensity is real.

Hold the photon flux, not the distance. When the goal is an intensity dependence, the controlled variable is irradiance. Set 20, 50, 100 and 200 mW/cm2 and adjust distance or neutral-density filtering to hit those values; the distances will come out uneven, and that is correct.

Watch the thermal side. Closing from 300 mm to 100 mm multiplies the IR load too, and a window that sat at 40 °C can reach 90 °C. On a photothermal catalyst this is not a nuisance, it is the experiment: separate the photon effect from the temperature effect by measuring the bed temperature in each position and reporting it with the rate.

Check meter linearity. Silicon photodiodes saturate or switch ranges near 100 mW/cm2, exactly where close-distance measurements land. Verify the reading against a thermal sensor or against a second distance where both are on-scale.

Frequently asked questions

Does irradiance really follow the inverse-square law in a photoreactor?

Only approximately, and never with the exponent 2 measured from the arc. A reflector and a finite arc shift the effective origin 40 to 70 mm behind the lamp, so the apparent exponent over 100 to 600 mm is closer to 1.6. Fit the exponent from your own measurements instead of assuming 2.

Where should the lamp-to-reactor distance be measured from?

From the arc centre to the outer face of the reactor window, with the window fitted and the beam normal to it. State this convention in the methods section, because the 30 to 50 mm between the housing face and the window is worth 30 % of the irradiance at short distances.

How much does moving the lamp closer increase the reaction rate?

Less than the irradiance increase, if the reaction is already light-saturated. Above roughly 100 mW/cm2 a catalyst limited by charge separation gains only a few percent per doubling, while below 20 mW/cm2 the rate often scales linearly with irradiance. Measure the rate at four irradiance levels rather than assuming either regime.

Why does my irradiance drop faster than expected at long distance?

Because the beam diverges past the reflector's design cone and the aperture starts to vignette. Beyond about 400 mm the useful spot also grows, so the on-axis value falls while the total power falls more slowly. Measure at the plane where the reactor actually sits, not at the focal point.