A 4.5-grade CO2 cylinder (99.995 %) may carry 3–5 ppmv of hydrocarbons quoted as methane. In a 1.0 L headspace at 25 °C and 1 atm that is 0.13–0.22 µmol of CH4 before the lamp is switched on — 10–50 % of a typical 4 h run. Grade 5.0 (99.999 %) cuts the same blank to 0.02–0.045 µmol.
A grade number describes total purity, not individual hydrocarbons. "4.5" means 99.995 % CO2; the remaining 50 ppmv is dominated by N2, O2, H2O and CO, and only a fraction of it is hydrocarbon. Because the total is what the grade certifies, a cylinder can pass 99.995 % while carrying 5 ppmv CH4. Suppliers differ too: some certificate 4.5 grade at ≤ 4 ppmv total hydrocarbons as CH4, others at ≤ 10 ppmv, so read the analysis certificate rather than the label.
The cylinder is rarely the only source. A second cylinder or hydrogen generator feeding the same reactor has its own methane spec. The GC carrier is a third: 99.999 % helium without a hydrocarbon trap still delivers a small CH4 peak at every injection, at a fixed retention time, looking exactly like product. Regulator diaphragms, septum bleed and new PTFE tubing add organics for the first few runs.
Convert ppmv to moles first: at 25 °C and 1 atm, 1 ppmv in 1 L is 1 µL, or 0.045 µmol. A 250 mL batch reactor at 5 ppmv therefore starts with 0.056 µmol CH4, a 1 L headspace with 0.22 µmol, and a 5 L gas bag used in flow mode with 1.1 µmol. Pressurising to 0.2 MPa gauge triples each figure, since amount scales with absolute pressure.
| Cylinder grade | Certified CO2 | Typical THC spec (as CH4) | CH4 in 1.0 L at 1 atm | Where it is acceptable |
|---|---|---|---|---|
| 3.0 | 99.9 % | ≤ 50 ppmv | 2.2 µmol | Not for trace CH4 quantification |
| 4.0 | 99.99 % | ≤ 10 ppmv | 0.45 µmol | Screening runs yielding > 20 µmol |
| 4.5 | 99.995 % | ≤ 3–5 ppmv | 0.13–0.22 µmol | Routine work with a measured blank correction |
| 5.0 | 99.999 % | ≤ 0.5–1 ppmv | 0.02–0.045 µmol | Yields of 0.1–1 µmol, most CO2 methanation |
| 5.5 | 99.9995 % | ≤ 0.1 ppmv | < 0.005 µmol | Isotope tracing and ppb-level detection |
The comparison that matters is blank versus yield in the same units. A 50 mg TiO2 sample giving 0.5 µmol CH4 in 4 h from a 4.5-grade cylinder in a 1 L headspace reports a signal in which 26–44 % of the carbon never saw a photon. Two labs with the same catalyst and different grades can differ twofold for that reason alone.
Fill the reactor through the same regulator, tubing and valve path as the experiment, at the same pressure and headspace, with catalyst present. Keep it dark for the full run and sample at identical intervals with the same loop. Three repetitions minimum: their standard deviation is the blank uncertainty, and 3σ of it the detection limit.
Add a carrier blank and an empty-reactor blank with no catalyst. If the three agree within 15 %, the cylinder dominates and a correction is defensible; if they differ by more than 30 %, the source is unstable — usually a regulator elastomer or a septum — and subtraction only hides it. Record the cylinder serial number; the blank changes with every fill lot.
Five checks separate them without new equipment. First, a light-off / light-on pair with identical gas charge: a cylinder blank is unchanged by illumination, a photocatalytic product is not. Second, an argon charge instead of CO2, irradiated — any CH4 that survives cannot come from CO2. Third, a catalyst-free irradiated control, which catches photolysis and chamber outgassing. Fourth, the time profile: a blank contributes a fixed amount per injection, whereas a product accumulates.
Fifth, isotope labelling. Run the experiment with 13CO2: genuine methanation gives 13CH4 at m/z 17, while cylinder methane remains 12CH4 at m/z 16. A signal that stays at m/z 16 is contamination, however linear the time plot looks. The same logic applies to CO, which is often present in the cylinder at 1–10 ppmv and is the more common false positive in CO2 reduction work.
Subtract absolute moles, never concentrations, and only when the blank is stable. Report corrected and uncorrected values, the cylinder grade, headspace volume and detection limit. A yield below three times the blank belongs in the report as "below blank + LOD", not as a number with two significant figures.
If 4.5 grade is what the lab has, fit an activated-carbon or molecular-sieve hydrocarbon trap on the CO2 line and verify it with a blank. A saturated trap passes methane unchanged and gives no warning.
No. The grade certifies total purity; hydrocarbon is one line of the certificate. Two cylinders of the same grade from different fill lots can differ threefold in THC, and fittings downstream can add more methane than the cylinder holds. Measure the blank on every new cylinder.
Only if the blank is reproducible. Require a relative standard deviation below 15 % over three runs, subtract in absolute micromoles, and report the uncorrected value as well. If the blank scatters by more than 30 %, find and remove the source instead of averaging it away.
Certificates for 4.5 grade commonly allow 3–5 ppmv total hydrocarbons as CH4, i.e. 0.13–0.22 µmol per litre of headspace. That is invisible in a 100 µmol demonstration run and dominant in a 0.5 µmol mechanistic study.
For yields above roughly 5 µmol CH4, yes, provided you measure and subtract the blank. Below 1 µmol, switch to 5.0 grade or install a hydrocarbon trap; otherwise the blank will dominate the error budget and no amount of replicate runs will recover the true rate.