Before any sophisticated spectroscopy, there is a simple, decisive question: does current change when the light turns on? Transient photocurrent and photovoltage measurements answer it, and the shape of the answer reveals more than the yes itself.
Illuminate the sample under bias with a chopped beam and record the current or photovoltage over many on-off cycles. The rise on illumination is the photoresponse; the decay after cutoff shows relaxation. The experiment costs minutes and diagnoses the photoelectric behavior comprehensively.
A sharp rise, flat plateau and sharp fall indicate clean photoresponse with fast carriers. A slow rise suggests trap filling or slow equilibration. A spike-and-decay on illumination, overshoot followed by sagging, indicates recombination competing with transport, a classic signature of poor charge separation. Anodic versus cathodic transient direction encodes carrier type and field direction.
Stable light source with clean switching (mechanical choppers or fast shutters); bias held constant; and recording bandwidth fast enough for the transient features, not just the steady state. Normalize by light intensity and sample area, and always run dark controls to verify the switching itself does not inject artifacts.
Steady photocurrent alone cannot distinguish a material that reaches its plateau instantly from one that spends minutes crawling there, but device behavior and photocatalysis absolutely depend on that difference. Transients expose the dynamics that steady measurements average away.